Balances are used to weigh different things as raw material, dispensed material and samples. Sometimes very low quantity is weighed for analysis. An error in the weighed material can cause a big variation in analytical results.
Following circumstances may cause the error in the weighing:
Assay, related substances and other quantitative analytical tests require accurate weighing for better results. Near the lower and upper capacity of the balance, weighing may be inaccurate. Therefore, a weighing range must be defined for all balances.
Related: Calibration of Analytical Balance
If any balance has the operating range from 0.01 mg to 110 gm then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic Weighing Instruments.
But according to USP <41> Weights and Balances, a simple theoretical calculation can be used to calculate the minimum weight for any balance. To weigh 10 mg sample you have to use 10X0.1% = 0.01 mg least count balance. It shows that if you have a balance with least count 0.01 mg, you can weigh minimum 10 mg sample on it. Other weights can be calculated accordingly.
Update: Simply you can calculate the minimum limit by multiplying least count by 1000.
Minimum Weighing Limit = Least Count x 1000
- Balance for analytical use must be closed; an open balance may cause inaccurate weighing.
- Hygroscopic material gains weight when it comes in contact with the air causing unstable weighing.
- Airflow or current in the balance room may cause a disturbance in accurate weighing.
- Unleveled surface or balance may lead to inaccurate weighing.
- A vibrating surface can also cause inaccurate weighing.

Related: Calibration of Analytical Balance
Weighing Range of Balances:
Some manufacturers direct in their operating manual to weigh minimum 1.0 mg on balance having 0.01 mg least count i.e. least count X 100. But the thumb rule followed in pharmaceutical industries for lower limit is least count X 50 and the upper limit is 80% of the capacity of the balance.If any balance has the operating range from 0.01 mg to 110 gm then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic Weighing Instruments.
Update: Simply you can calculate the minimum limit by multiplying least count by 1000.
Minimum Weighing Limit = Least Count x 1000
please give us the ISO and Auditing documents of electrical dept also sir...
ReplyDeleteinformative article
ReplyDeleteWhere to find the least count for a specific weighing balance
ReplyDeleteNo need to perform daily verification because all ranges covered in monthly calibration
ReplyDelete