Relative Response Factor (RRF) and its Calculation in HPLC Analysis : Pharmaguideline
  •   Pharmaguideline Courses Enroll Now
  • Apr 22, 2024

    Relative Response Factor (RRF) and its Calculation in HPLC Analysis

    Know about the relative response factor and its calculation in system suitability parameters in HPLC analysis of pharmaceutical substances and pharmaceutical products.
    During the manufacturing process of active pharmaceutical ingredients, some unwanted substances are produced those are known as impurities. Some impurities are raised due to the degradation of the actual drug.

    Impurities in pharmaceutical drug substances and drug products are required to be analyzed for the quantity of the impurities. To determine the actual quantity of the impurity it is required to have impurity standard. In the absence of the impurity standard relative response factor (RRF) may be used to calculate the actual quantity of the impurity.

    Determaination of Response Factor
    Q3B R2 of ICH Guidelines also says that if the response factor is determined correctly, it can be used to measure the actual amount of impurity. Identification threshold, quantification threshold and reporting threshold are given by ICH and it is important to determine the impurities within these thresholds.


    Relative Response Factor (full form of RRF) is an alternate method for the determination of the quantity of the impurities present in pharmaceutical products and amount of the impurity can be calculated with the help of peak area of the components. 

    Relative response factor is the ratio of the response of the impurity and the active pharmaceutical ingredient (API) under the identical chromatographic conditions (chromatographic column, temperature, mobile phase, flow rate etc). 

    Relative response factor is determined by analyzing the impurity standard and API standard of equal concentration. The following formula is used to determine the response factor:

    Response Factor (RF) =           Peak Area         
                                        Concentration in mg/ml

    Relative Response Factor (RRF)  =  Response Factor of impurity
                                                              Response Factor of API

    RF in chromatography for different products are different and should be determined for individual substance. Relative response factor determined on a particular detector cannot be used in the calculation of product determined on another detector because different detectors show the difference in the response of the same compound.

    Solutions of at least three different concentrations of standard and impurity are prepared and injected in HPLC. The slop of area and concentration of impurity and standard are calculated to determine the relative response factor.

    1 comment: Post Yours! Read Comment Policy ▼

    1. Need to mentioned about the reciprocal or direct multiplication conditions at the time of final calculation.

      ReplyDelete

    Please don't spam. Comments having links would not be published.


    Popular Categories

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses


    Follow Pharmaguideline



    DOCUMENTS

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App