1.0 Objective
To calibrate the equipment for reliable and accurate results.2.0 Scope
This procedure is applicable to calibrate the Karl Fischer apparatus, installed in the quality control department.3.0 Responsibility
3.1 Doing: Technical Assistant3.2 Checking: Executive/Asst. Manager
4.0 Accountability
Head of the Department5.0 Procedure
Frequency: Once in a month5.1 Operate the instrument as per SOP.
5.2 FOR BURETTE CALIBRATION
5.2.1 Place the burette outlet in 20ml standard calibrated measuring cylinder.5.2.2 Dispense 2,4,6,8,10,12,14,16,18 & 20 ml water into previously weighed & conditioned beaker and note the water weight of water for the dispensed.
5.2.3 Note down the observed volume of water in the measuring cylinder.
5.2.4 Record the obtained volume and displayed volume in the enclosed format.
5.2.5 The volume displayed on the instrument and the volume collected in cylinder should not vary by more than 0.5 %, otherwise, follow the S.O.P.
5.2.6 Record the results in Annexure.
5.2.7 Frequency of calibration: Once in a month.
5.3 FOR INSTRUMENT PERFORMANCE
5.3.1 Add accurately weighed 50mg, 100mg, 150mg, 200mg, 250mg of Disodium tartrate into the titration vessel containing neutralized methanol and test for water content.5.3.2 Calculate the % water by using the formula given below.
Volume of KFR consumed in ml X 100 X Factor
% Water (W/W) = --------------------------------------------------------------
Weight of disodium tartrate in mg
5.3.4 Record the results in Annexure-I.
5.3.5 Frequency of calibration: Once in a month.
6.0 ABBREVIATIONS
6.1 % = Percentage6.2 mg = milligram
6.3 W/W = Weight by Weight
6.4 S.O.P. = Standard operating procedure
6.5 ml = milliliter
6.6 KFR = Karl fischer reagent
ANNEXURE-I
Q. C. Department
PERFORMANCE CHECK REPORT OF KARL FISCHER APPARATUS
PERFORMANCE CHECK DATE
|
DATE OF LAST PERFORMANCE
CHECK DONE
|
NEXT DUE FOR
PERFORMANCE CHECK
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
|
INSTRUMENT MAKE
|
INSTRUMENT ID No.
|
OBSERVATION |
Sr. No.
|
Wt. taken of di-sodium tartrate
(mg)
|
Volume of KFR required
(ml)
|
Factor
mg. H2O/ml
|
% Water obtained
|
Tolerance
|
1
|
14.87 % to 16.44 %
| ||||
2
| |||||
3
| |||||
4
| |||||
5
| |||||
Linearity : ____(NLT 99.00%)
|
% RSD :
|
(NMT 1.0 %)
|
Calculation :
BURETTE CALIBRATION :
Actual temp of water : ________
Factor : ________
Wt. of beaker (A) : _________
Sr. no.
|
Displayed volume
(ml)
B
|
Wt. of beaker + water
(gm)
C
|
Wt. of water
(gm)
D=(C-A) gm
|
Actual volume (ml)
E=
(Factor X D)
|
difference
F=(E-B)
|
Tolerance + 0.5 % of displayed volume
|
1
| ||||||
2
| ||||||
3
| ||||||
4
| ||||||
5
| ||||||
6
| ||||||
7
| ||||||
8
| ||||||
9
| ||||||
10
|
Calculation :