1.0 Objective
To check the performance of Autodiluter.2.0 Scope
This procedure is applicable to calibrate Autodiluter.3.0 Responsibility
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 Accountability
Head of the Department5.0 Procedure
5.1 Frequency: once in a 3 months5.1 Clean the instrument with a dry cotton cloth.
5.2 Operate the instrument as per S.O.P.
5.3 For Right syringe (B)
5.3.1 Take five empty beaker, they should be conditioned & weigh previously.
5.3.2 Suck 100µl with the right syringe (B) & Note down the water weight. Further take 200µl, 300µl, 400µl & 500µl & Note down the water weight.
5.3.4 Check the linearity as shown in Annexure-I.
5.4 For left syringe (A)
5.4.1 Take five empty beaker, they should be conditioned & weigh previously.
5.4.2 Suck 1 ml with the left syringe (A) & note down the water weight.
5.4.3 Further take 2 ml, 3 ml, 4 ml, & 5 ml and note down the weight.
5.4.4 Check the linearity as shown in Annexure –II.
6.0 ABBREVIATION
6.1 μl = Microliter6.2 ml = Millilitre
ANNEXURE –I
Q. C. Department
PERFORMANCE CHECK RECORD FOR AUTODILUTER
Name of Instrument : | Ref No. : |
Make : |
Issue Date :
|
Model No. : | Issued By : |
Id. No. : |
For Right Syringe (B)
Sr. No.
|
Volume
|
Wt. Of water in mg
|
Linearity
|
1
|
100 µl
| ||
2
|
200 µl
| ||
3
|
300 µl
| ||
4
|
400 µl
| ||
5
|
500 µl
|
Linearity : Limit : NLT 99.0%
Remarks :
Calibration date :
Calibration due date :
Calibrated by :
Checked by :
ANNEXURE –II
Q. C. Department
For Left Syringe (A)
Sr. No.
|
Volume
|
Wt. Of water in mg
|
Linearity
|
1
|
1 ml
| ||
2
|
2 ml
| ||
3
|
3 ml
| ||
4
|
4 ml
| ||
5
|
5 ml
|
Linearity : Limit : NLT 99.0%
Remarks :
Calibration date :
Calibration due date :
Calibrated by :
Checked by :