1.0 OBJECTIVE
To ensure constant temperature range in respective walk-in incubators.
2.0 SCOPE
This procedure is applicable to calibrate walk-in – incubators of 20-25°C and 30-35°C using Calibrated thermometer.
3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant (Q.C) /Executive
3.2 Checking: Executive /Manager
3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the department.
5.0 PROCEDURE
5.1 Frequency: Quarterly
Note: Calibrate the sensor of both walk .in –incubator by outside party at every six months.
5.2 Place four calibrated thermometers in the walk-in incubator as per location chart attached.
5.3 Take the readings from each thermometer at an interval of every two hours and record the results up to 8 hours in Annexure.
Note: Calibrate the sensor of both walk .in –incubator by outside party at every six months.
5.2 Place four calibrated thermometers in the walk-in incubator as per location chart attached.
5.3 Take the readings from each thermometer at an interval of every two hours and record the results up to 8 hours in Annexure.
6.0 ABBREVIATIONS
6.1 SOP - Standard Operating Procedure
ANNEXURE - I
MICROBIOLOGY LAB. Q.C.
CALIBRATION OF WALK – IN – INCUBATORS 20 – 25°C AND 30 – 35°C
(A) 20-25°C Date:
Thermo
ID. No.
|
Time
|
Location
| |||
A
|
B
|
C
|
D
| ||
Set temp __________
Acceptance criteria ( Set temp + 2° C
|
Date:
Thermo
ID. No.
|
Time
|
Location
| |||
A
|
B
|
C
|
D
| ||
Set temp __________
Acceptance criteria ( Set temp + 2° C
|
Remarks
Standard thermometers used
Calibration Date
Calibration Due Date
Calibration By
Checked By
|
: Satisfactory / not satisfactory
:
:
:
: : |