SOP for Calibration of pH Meter Temperature Sensor

Standard operating procedure for the calibration of pH meter temperature sensor using water for injection (WFI) / purified water (PW).

1.0 Objective

To provide a standard operating procedure for the calibration of pH meter temperature sensor.

2.0 Scope

This procedure is applicable to pH meter.

3.0 Responsibility

3.1 Doing: Technical Assistant/Executive
3.2 Checking: Executive /Manager

4.0 Accountability

Head of the Department

5.0 Procedure

Frequency: Twice in Year.
5.1 Calibration Procedure for pH meter temperature sensor
5.1.1 Keep the sensor probe of reference calibrated digital thermometer in the beaker containing water for injection (WFI) / purified water (PW) of about temperature 35°C.
5.1.2 Add cold WFI/PW slowly drop wise and bring the temp. between 27°C to 30°C.
5.1.3 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.4 Add WFI/PW in some more quantity slowly drop wise to bring the temp. between 26°C to 23°C.
5.1.5 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.6 Add cold WFI/PW slowly drop wise and bring the temp. between 20°C to 22°C.
5.1.7 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.8 Acceptance Criteria : ± 3°C

6.0 Abbreviations

6.1 °C= Degree centigrade

ANNEXURE-I
Q. C. Department
PH METER TEMPERATURE SENSOR RECORDS
                                                                   
REF.No. :_________________
ISSUE DATE: _________________
ISSUED BY :____________________   



Calibrated  Ref. Digital Thermometer No.:____________________ 
Ref. Thermometer Calibration date          :____________________
Ref. Thermometer Certificate No.            :____________________
Ref. Thermometer Calibration Due Date  : ____________________
pH meter ID.No.                                      : ___________________
Results Observed:
Set Range
Observation
Difference
tolerance
Temperature as read on pH meter
Temperature as on read on display of Ref. Thermo.
30°C to 27°C



±0.3°C
26°C to 23°C



22°C to 20°C




Remarks                 :
Calibration date          :
Next Calibration date  :
Calibrated  by             :
Checked by                 :
Get ready to use editable documents in MS-Word FormatView List





Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


No comments:

Post a Comment

Please don't spam. Comments having links would not be published.


Popular Categories

QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
Online Courses


Follow Pharmaguideline


DOCUMENTS

PHARMACEUTICAL DOCUMENTS




Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

View


adsbypg


Recent Posts