SOP for Performance Check of Fourier Transform Infrared : Pharmaguideline

Online GMP Courses with Certificate

ENROLL

SOP for Performance Check of Fourier Transform Infrared

Standard operating procedure to evaluate the performance of Fourier Transform Infrared Spectrophotometer using Polystyrene Film.

1.0 Objective

To check the performance of Fourier Transform Infrared Spectrophotometer.

2.0 Scope

This procedure is applicable to calibrate Fourier Transform Infrared Spectrophotometer.

3.0 Responsibility

3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager

4.0 Accountability

Head of the Department

5.0 Procedure

5.1 Clean the instrument with dry cotton cloth.
5.2 Switch 'ON' the instrument.
5.3 Switch 'ON' the computer.
5.4 Click on 'cancel' at network password.
5.5 Click on 'cancel' at Update Device Driver Wizard.
5.6 Double-click on 'Shimadzu FTIR -8900'.
5.7 Click on O.K. at Shimadzu Hyper IR.
5.8 “Online FTI” will appear on lower left side of the screen.
5.9 Place sample holder in the chamber.
5.10 Click 1
5.11 “Which spectra do you choose”shall be appeared.
5.12 Click on ‘measure’.
5.13 Message “Remove sample from holder” shall be appeared.
5.14 Click “O.K.”
5.15 “Shimadzu FTIR 8000”default shall be appeared.
5.16 “Background graph” shall be appeared.
5.17 Message “Set polystyrene film” shall be appeared.
5.18 Place polystyrene film in the holder.
5.19 Click ‘O.K’
5.20 “Shimadzu FTIR 8000 default ” shall be appeared.
5.21 IR appears on the screen
5.22 Results appears on the screen.
5.23 Message’ ‘judgement O.K.’ shall be appeared.
5.24 Message ‘output printer’ shall be appeared.
5.25 Click ‘yes’.
5.26 Check following parameters from the print out
5.26.1 Power spectrum
5.26.2 Resolution
5.26.3 Wave number accuracy
5.26.4 Repeatability of wave number
5.26.5 Repeatability of Transmittance
5.27 Enter the results of the above parameters in Annexure

6.0 Abbreviations

6.1 I.R = Infrared
6.2 FTIR = Fourier transform infrared

Annexure-I
Q. C. Department
 PERFORMANCE CHECK RECORD OF
FOURIER TRANSFORM INFRARED SPECTROPHOTOMETER
Name of the Instrument :  FOURIER TRANSFORM INFRARED SPECTROPHOTOMETER  
Make                                             : 
Model No                                   : 
IDENTIFICATION NO.               :
*  Power Spectrum :
   Wavenumber
Measured
Intensity specification
4600

10% or more of maximum
4000

30% or more of maximum
3000

60% or more of maximum
At maximum

60 or more
700.0

10% or more of maximum
500.0

2% or more of maximum
403.0

0.5% or more of maximum

   Remarks : Satisfactory/Not  Satisfactory
*  Resolution
   Wavenumber
Measured (T%)
Standard peak depth (%T)
2870, 2851, 2868.9  2848.7

Not less than 18
1589.0 , 1588.3
1583.0 , 1582.5

Not less than 10

   Remarks : Satisfactory/Not  Satisfactory
 *  Wave number accuracy
Wavenumber
Measured wave number
Error
Tolerance
3027.1


+/- 5.0
2850.7


+/- 5.0
1601.4


+/- 1.0
1583.1


+/- 1.0
1181.4


+/- 1.0
1069.1


+/- 1.0
906.7


+/- 1.0


  Remarks : Satisfactory/Not  Satisfactory


                                                                                                                         Page 2 of 2



Name of the Instrument   : FOURIER TRANSFORM INFRARED SPECTROPHOTOMETER  

Make                                               : 

Model No                                     : 
IDENTIFICATION NO.                 :
--------------------------------------------------------------------------------------------------------------------------------- 
* Repeatability of wave number :
Wavenumber
Measured- I
wave number
Measured  -II
wave number
Error
Tolerance
2850.7



+/- 5.0
1601.4



+/- 1.0
1181.4



+/- 1.0
   Remarks : Satisfactory/Not  Satisfactory
* Repeatability of Transmittance :
Wavenumber
Observation -I
        %T
Observation-II   
        %T
Error
Tolerance
2850.7



+/- 0.5
1601.4



+/- 0.5
1181.4



+/- 0.5

     Remarks                       : Satisfactory/Not Satisfactory
     Performance check  done by          :                                                     Date :    
     Checked by                                     :                                                     Date :
     Next Performance check  due date :
Get ready to use editable documents in MS-Word FormatView List





Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


No comments:

Post a Comment

Please don't spam. Comments having links would not be published.


Popular Categories

QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
Online Courses


Follow Pharmaguideline


DOCUMENTS

PHARMACEUTICAL DOCUMENTS




Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

View


adsbypg


Recent Posts