1.0 Objective
To calibrate the equipment for confirmation of reliable and accurate results.2.0 Scope
This procedure is applicable to calibrate the refractometer, installed in the quality control department.3.0 Responsibility
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 Accountability
Head of the Department5.0 Procedure
Frequency: Once in three months.5.1 Operate the instrument as per S.O.P.
5.2 Maintain the required temperature of the particular liquid as mentioned in Annexure-I
5.3 Observe and record refractive index of the different liquid at temperature mentioned in Annexure-I.
5.4 The performance of the instrument is satisfactory if the observed refractive index is within the limits given in Annexure, otherwise follow the S.O.P.
6.0 Abbreviation
6.1 SOP - Standard Operating ProcedureANNEXURE-I
QUALITY CONTROL LABORATORY
CALIBRATION REPORT OF REFRACTOMETER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
Sr.No.
|
Liquid
|
Temperature
|
Observed Refractive index
|
Limit
|
01
|
Distilled water
|
20oC
|
1.330+0.0001
| |
02
|
Carbon tetrachloride (AR grade)
|
20oC
|
1.459 – 1.461
| |
03
|
Toluene (AR grade)
|
20oC
|
1.495-1.497
|