1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.2.0 SCOPE
This procedure is applicable to calibrate Head Space Gas Chromatograph installed in Quality Control Laboratory.3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant3.2 Checking: Executive/Manager
4.0 ACCOUNTABILITY
Head of the Department5.0 PROCEDURE
5.1 (I) BY MANUAL INJECTION:
Frequency: Every 3 month.
5.1.1 Install the DB-5 capillary column.5.1.2 Set all parameters as per Annexure – I.
5.1.3 Stabilized the column for the stable baseline.
5.1.4 Inject 1 µl of each standard solution in duplicate.
5.1.5 Calculate the ratio of Area of methanol to the internal standard.
5.1.6 Record the results in Annexure – I.
5.1.7 Calculate the linearity and correlation coefficient.
5.2 (II) FOR HEADSPACE INJECTION :
Frequency: Every 6 months.
5.2.1 Install the DB-5 capillary column.5.2.2 Set all parameter as per Annexure – II
5.2.3 Prepare the standard solution as mentioned in Annexure – II
5.3 Check the performance for
(1) Reproducibility test
(2) Linearity test.5.4 Reproducibility test
5.4.1 Take 5 ml of standard preparation in 5 Head Space vial.5.4.2 Add 1 gm of sodium sulphate (Anhydrous) in each.
5.4.3 Inject 1 ml gaseous form of the standard from each of 5 vials.
5.4.4 Record the result in Annexure – II
5.4.5 Calculate the standard deviation and related standard deviation for the respective separated solvent peak.
Related: Calibration of Gas Chromatography (GC)
5.5 Linearity test
5.5.1 Take 2 ml, 5 ml and 10 ml of standard solution in Head Space Vial.5.5.2 Add 1gm sodium sulphate (Anhydrous) make 10 ml volume by addition of required volume in respective headspace vial.
5.5.3 Inject 1 ml Gaseous form of each standard.
5.5.4 Record the result in Annexure – II
5.5.5 Calculate the regression coefficient of the respective separated solvent peak.
5.6 After completion of Analysis, cool down the G.C.
5.7 After cooling, switch ‘OFF’ the instrument.
5.8 Switch ‘OFF’ the mains.
6.0 ABBREVIATIONS
6.1 G.C. = Gas chromatograph6.2 ml = millimeter
ANNEXURE-I
QUALITY CONTROL LABORATORY
CALIBRATION CHECK REPORT OF
GAS CHROMATOGRAPH
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
(I) FOR MANUAL INJECTION
G.C. Parameters:
Column : DB-5 Capillary (0.32mm X 30meter X 0.25micron)
Range : 2
Oven Temp. : 300C
Split Ratio : 1:20
Detector Temp : 1800C
Nitrogen : 100 kPa
Inj. Temp. : 1400C
Carrier : 25Kpa (Helium)
Inj. Volume : 1.0μl
Hydrogen : 55Kpa
Air : 85 Kpa
G.C. Parameters:
Column : DB-5 Capillary (0.32mm X 30meter X 0.25micron)
Range : 2
Oven Temp. : 300C
Detector Temp : 1800C
Nitrogen : 100 kPa
Inj. Temp. : 1400C
Carrier : 25Kpa (Helium)
Inj. Volume : 1.0μl
Hydrogen : 55Kpa
Air : 85 Kpa
Preparation of internal Standard: 25 ml Ethanol è 1000 ml with water
Test preparation : 1 ml, 2 ml, 3 ml, 4 ml, 5 ml Methanol 100 ml with Internal Std.
Test preparation : 1 ml, 2 ml, 3 ml, 4 ml, 5 ml Methanol 100 ml with Internal Std.
Methanol (% v/v) |
Ratio of Methanol/Ethanol
|
Mean Ratio
|
1.0
|
(1)
| |
(2)
| ||
2.0
|
(1)
| |
(2)
| ||
3.0
|
(1)
| |
(2)
| ||
4.0
|
(1)
| |
(2)
| ||
5.0
|
(1)
| |
(2)
|
Correlation coefficient :___________ (Not less than 0.9900)
PERFOMED BY | CHECKED BY | ||||
NAME
|
SIGN.
|
DATE
|
NAME
|
SIGN.
|
DATE
|