1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.2.0 SCOPE
This procedure is applicable to calibrate Hardness tester, installed in the quality control area.3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant / Executive3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the department.5.0 PROCEDURE
5.1 Select “configuration” menu from the main menu.5.2 Select “calibration” sub menu and press enter
5.3 Enter password “Calibration” and press enter.
Following four matters can be calibrated by this instrument.
1) Balance
2) Hardness tester
3) Diameter
4) Thickness
5.4 Calibration of Mettler Balance
5.4.1 Select the menu ‘Balance’ and press enter.5.4.2 After a first beep tone, a weight value at 50.000 gm is indicated by blinking.
5.4.3 Put 50.000 gm weight onto the balance boat and close the cover.
5.4.4 After a second beep tone give the massage of remove 50.000 gm weight from the balance boat.
5.4.5 Confirm the calibration with ‘enter’ and check the accuracy of the balance with 3 different calibrator weight and enter as nominal value and receive actual value.
5.6 Calibration of Diameter
5.6.1 Select the menu item ‘Diameter’ from the sub menu ‘calibration’ and enter.5.6.2 A gauge block size of 10 nm is entered as standard.
5.6.3 To check the accuracy of Diameter calibration with 3 different calibrated gauge block.
5.6.4 Enter the nominal value and receive actual value.
Related: SOP for Cleaning of Tablet Hardness Tester
5.7 Calibration of Thickness
5.7.1 Select the menu item ‘Thickness’ from the sub menu ‘calibration’ and enter.5.7.2 Clean the measuring plate and the tablet track with brush and press enter.
5.7.3 Insert 5mm calibrated block and press enter.
5.7.4 To check the accuracy of thickness calibration with 3 different calibrated gauge block.
5.7.5 Enter the nominal value and receive actual value.
5.7.6 After exit the calibration menu, calibration protocol printout by ‘Print out OK’ message.
5.8 Frequency: Once in 3 month
6.0 ABBREVIATION
6.1 SOP - Standard Operating Procedure
ANNEXURE-I
QUALITY CONTROL LABORATORY
CALIBRATION REPORT OF HARDNESS TESTER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
CALIBRATION PROTOCOL
1) For mattler Balance :
(Last calibration with _____ gm)
Sr. No.
|
Wt. In gm
|
Limit + 0.5%
of Nominal value
| ||
1
|
0.5000
|
Nominal
| ||
Actual
| ||||
2
|
5.000
|
Nominal
| ||
Actual
| ||||
3
|
50.000
|
Nominal
| ||
Actual
|
2) For Hardness tester :
(Last calibration with _____ Kg)
Sr. No.
|
Wt. In Kg
|
Limit + 2.0%
of Nominal value
| ||
1
|
5
|
Nominal
| ||
Actual
| ||||
2
|
10
|
Nominal
| ||
Actual
| ||||
3
|
20
|
Nominal
| ||
Actual
|
OBSERVATION
|
CALIBRATION PROTOCOL
3) For Diameter :
(Last calibration with _____ mm)
Sr. No.
|
Diameter in mm
|
Limit + 0.05 mm
of Nominal value
| ||
1
|
5.00
|
Nominal
| ||
Actual
| ||||
2
|
10.00
|
Nominal
| ||
Actual
| ||||
3
|
15.00
|
Nominal
| ||
Actual
|
4) For Thickness :
(Last calibration with _____ mm)
Sr. No.
|
Thickness in nm
|
Limit + 0.05 nm
of Nominal value
| ||
1
|
3
|
Nominal
| ||
Actual
| ||||
2
|
5
|
Nominal
| ||
Actual
| ||||
3
|
8
|
Nominal
| ||
Actual
|
Get ready to use editable documents in MS-Word FormatView List