1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.2.0 SCOPE
This procedure is applicable to calibrate the coulter counter installed in the quality control department.3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the Department5.0 PROCEDURE
Frequency: Once in a three month5.1 Program and operate the instrument as per SOP.
5.2 Use the ‘menu cursor’ keys to select the following option on the ‘SET UP’ menu.
5.3 Set the parameters as follows.
Orifice size :70 & 100 µ alternately
Diameter/Length: Enter the aperture diameter and length of the orifice tube selected for calibration.
Set up: Select 'Automatic'
Analysis: Select 'Sample'
Calibration: Select 'New'
Size: Enter the (number mode) of the latex particles used for calibration and the assay Diameter of the chosen latex size should be 20 % of the aperture diameter for optimum results.(e.g. 2.15 E1 means 21.5 m)
Size units: Select mm
5.4 Select ‘FULL RANGE’.
5.4.1 X-Axis ----> Select ‘Lin Dia’.
5.4.2 Y-Axis ----> Select ‘Number Diff’.
5.4.3 Blank subtract ---> Select No.
5.4.4 Channels-----> 256
5.4.5 Channel data --> No.
5.5 Observe reading as per S.O.P.
5.5.1 Move the upper and lower cursor to coincide with peak height.
5.5.2 Press SET UP key and select record on the calibration line on SET UP menu.
5.5.3 The new Kd factor will display .
5.5.4 Press the CAL key, the instrument is now calibrated for that electrolyte solution and orifice tube for diameter and volume.
5.5.5 Print the graph and set up menu with new Kd factor for recording. Record the observation as per annnexure. If calibration is out of limit, follow the S.O.P.
6.0 ABBREVIATIONS
6.1 No.= Number6.2 Lin.Dia.= Linear diameter
6.3 Diff = Difference
6.4 mm= Micrometer
6.5 m= Micron
ANNEXURE-I
CALIBRATION REPORT OF COULTER COUNTER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
|
||
INSTRUMENT NAME
|
INSTRUMENT MAKE
|
INSTRUMENT
IDENTIFICATION NO.
|
Name of std. Calibration solution : CC standard L5 Polystyrene Latex
Lot No. :
Size of Latex particles :
Conductivity fluid : Filtered N.Saline
Label claim :
For 70µm aperture diameter : 3 drops per 50ml
For 100µm aperture diameter : 3 drops per 50ml
OBSERVATION
|
Orifice Diameter Observed Kd value Limit of Kd value
70 microns 677+/-10%
100 microns 930 +/- 10%
70 microns +/- 2 % of label claim. (Limit: to µm)