1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.2.0 SCOPE
This procedure is applicable to the infrared spectrophotometer, installed in the quality control department.3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the Department5.0 PROCEDURE
Frequency of calibration: Once in a month5.1 Operate the instrument as per SOP.
5.2 Put the reference polystyrene film (0.05 mm thick) in the sample holder and scan.
5.3 Compare the obtained spectrum with the reference spectrum for the following wave numbers as well as transmittance at shown wave number. Record the results as per annexure I.
5.3.1 Sr. No. Wave number % T
01 1583 cm-1 to 1589 cm-1 greater than 12%
02 2851 cm-1 to 2870 cm-1 greater than 18 %
5.3.2 Sr. No. Wave numbers of Reference Polystyrene film [Limit + 5 cm1 ]
1 3027.1
2 2924.0
3 2850.7
4 1944.0
5 1871.0
6 1801.6
7 1601.4
8 1583.1
9 1181.4
10 1154.3
11 1069.1
12 1028.0
13 906.7
14 698.9
5.4 The performance of the instrument is satisfactory, if obtained I.R. Spectrum shows peaks at above wave numbers, otherwise follow the S.O.P.6.0 ABBREVIATIONS
6.1 % = Percentage6.2 I.R = Infra red
6.3 S.O.P. = Standard operating procedure
ANNEXURE-I
QUALITY CONTROL LABORATORY
CALIBRTAION REPORT OF INFRARED SPECTROPHOTOMETER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENTS DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION
|
OBSERVATION
|
For Transmittance
For wave numbers
|