1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.2.0 SCOPE
This procedure is applicable to the Brook field viscometer, installed in the quality control department.3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the Department5.0 PROCEDURE
Frequency of calibration: Once in three month5.1 Operate the instrument as per SOP
5.2 Calibrate the instrument with standard viscosity solution of 10, 100, 500, 5000 & 12,500 cps using respective spindles and record the results in Annexure-I
5.3 The performance of the instrument is satisfactory, if the measured viscosity is within the given tolerance limit, otherwise follow the S.O.P.
6.0 ABBREVIATION
6.1 0C = Centigrade6.2 S.O.P. = Standard operating procedure
6.3 cps = Centipoise
Annexure-I
CALIBRATION REPORT OF BROOK FIELD VISCOMETER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
|
INSTRUMENT MAKE
|
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
Sr. No.
|
Fluid of Labeled viscosity in cps
|
Certified Value in cps
|
Spindle
|
Speed factor at 250C (A)
|
Reading (B)
|
Viscocity in cps AXB
|
Acceptance criteria % Tolerance from certified viscosity
|
01
|
10
|
+/- 10%
|
|||||
02
|
100
|
+/- 7.5%
|
|||||
03
|
500
|
+/- 7.5%
|
|||||
04
|
5000
|
+/- 5.0%
|
|||||
05
|
12,500
|
+/- 5.0%
|