1.0 OBJECTIVE
To clearly lay down the procedure for cleaning.2.0 SCOPE
This procedure is applicable for ultrasonic bath, installed at quality control laboratory.3.0 RESPONSIBILITY
3.1 Doing : Technical Assistant3.2 Checking : Executive/Manager
4.0 ACCOUNTABILITY
Head of the Department5.0 PROCEDURE
Frequency: Once in a week5.1 Switch ‘OFF’ the Mains.
5.2 Remove the lid from the ultrasonic bath.
5.3 Wash the lid with purified water properly .
5.4 Remove the water from the ultrasonic bath and wash the ultrasonic bath with purified water properly and mop with a clean dry cotton cloth.
5.4 Fill purified water up to 3/4 th height of the bath.
5.6 Put the lid of the bath.
5.7 After completion of cleaning, record in the format.
Note: In between when water becomes dirty follow the same from 5.1 to 5.7
6.0 ABBREVIATION
6.1 SOP - Standard Operating Procedure