1.0 OBJECTIVE
To describe the procedure for operation and calibration of Bursting strength tester.2.0 SCOPE
This SOP is applicable for operation and calibration of Bursting strength tester.3.0 RESPONSIBILITY
Officer/ Executive - Quality Control4.0 ACCOUNTABILITY
Manager - Quality Control5.0 PROCEDURE
5.1 Operation
5.1.1 Ensure that the area and instrument is clean and the pressure gauges are within the due period of calibration.5.1.2 Switch “ON” the instrument by “ON/OFF” switch.
5.1.3 Turn the clamping wheel in the anticlockwise direction.
5.1.4 Take the sample to be tested and place it over the diaphragm plate.
5.1.5 Clamp the sample by applying adequate pressure by the clockwise turn of the wheel.
5.1.6 Select the gauge to be used depending on the sample to be tested. For lower value, like in case paper and low range board use the lower range gauge by turning the rotary switch to Gauge 2 position. In case of the heavy board, turn back the switch to Gauge 1 position.
5.1.7 Bring the lazy hand (a red indicator of the gauge) to zero position.
5.1.8 Press the “PUSH FOR BURST” button and keep it pressed till the sample burst.
5.1.9 Record the maximum value in kg/cm2 in Annexure- I
5.2 Calibration
5.2.1 Calibrate the Gauge meter by the external agency.5.2.2 Operate the instrument as per the operating procedure and check the bursting strength of standard aluminum foil, which bursts at 10 kg/cm2.
Std. Foil Bursting Strength
|
Observed Bursting Strength
|
Tolerance
|
(1) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
|
(2) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
|
(3) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
5.3 Precaution
5.3.1 Care has to be taken that the gauge is not used beyond their capacity as it may damage the gauge.5.3.2 It is a hydraulic unit. Its accuracy depends upon a totally filled hydraulic system. Fluid from the system may be lost by leaked or diaphragm breakage. It is replaced by air from the atmosphere. Such accumulated air in the system forms an air cushion, which absorbs pressure and affects the accuracy of the test results erratically.
6.0 ABBREVIATIONS
6.1 SOP - Standard Operating Procedure6.2 No. - Number
6.3 QA - Quality Assurance
6.4 QC - Quality Control
6.5 NA - Not Applicable
ANNEXURE-I
BURSTING STRENGTH TESTER USAGE LOG BOOK
Date
|
Product/
Material Name
|
Batch No./ Lot No.
|
Time
|
Analyzed
by
|
Checked by
|
Remarks
|
|
Started
at
|
Completed
on
|
||||||
ANNEXURE-II
CALIBRATION REPORT OF BURSTING STRENGTH TESTER
Asset ID No.
|
:
|
Ref. SOP No.
|
:
|
||
Make
|
:
|
Model No.
|
:
|
Std. Foil Bursting Strength
|
Observed Bursting Strength
|
Tolerance
|
(1) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
|
(2) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
|
(3) 10 Kg/cm2
|
± 0.5 Kg/cm2
|
Remarks: The instrument is calibrated as per the SOP and shows satisfactory / does not satisfactory results as per calibration procedure.
Calibrated on : ____________________
Next calibration due on : ____________________
Done by: Checked
Date: Date: