PROCEDURE
Calibrate the apparatus using following solvents:
Name of the Liquid
|
Refractive Index
|
Temperature
|
Distilled Water
|
1.3325
|
20°C ± 0.5°C
|
Toluene
|
1.4969
|
20°C ± 0.5°C
|
Carbon Tetrachloride
|
1.4603
|
20°C ± 0.5°C
|
Frequency: Once in a month.
Maintenance/ Repairs: When the instrument does not comply with the requirement/ tolerance range specified above, the instrument should be labeled “OUT OF CALIBRATION” and should be repaired/ serviced.
Reference: Operator’s instructions manual.
QUALITY ASSURANCE DEPARTMENT CALIBRATION REPORT OF REFRACTOMETER |
||||||||||||||||||||
Reference :
|
Date -
|
|||||||||||||||||||
Frequency -
|
Instrument No. -
|
|||||||||||||||||||
Make –
|
Model –
|
|||||||||||||||||||
|
||||||||||||||||||||
Conclusion :
Calibration complies/ does not comply as per the acceptance criteria.
|
||||||||||||||||||||
Next calibration due on :
|
||||||||||||||||||||
Done By :
Date :
|
Checked By :
Date :
|
|||||||||||||||||||