PROCEDURE
- Ensure that all the connections of the instrument are proper.
- Turn the viscometer ON.
- Take out the viscosity standard solutions in 500 ml beaker, adjust the temperature to 250C. Attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.
- Viscosity standards, speeds, spindles required & tolerance limits are as follows:
Sr.
No.
|
Viscosity of standard in cps
|
Actual viscosity of standard in cps
|
Spindle No.
|
Entry Code
|
RPM
|
Tolerance (in cps)
|
|
Minimum
|
Maximum
|
||||||
1.
|
5.00
|
4.70
|
ULA
|
00
|
60
|
4.55
|
4.85
|
2.
|
50.00
|
48.40
|
LV1
|
61
|
60
|
46.9
|
49.9
|
3.
|
5000.00
|
4820.00
|
LV3
|
63
|
10
|
4651.83
|
4988.17
|
Acceptance criteria
- Viscosity should be within the tolerance limits.
- Do not transfer the viscosity solution to the original container back after use.
Precautions
- Take due care to avoid any contamination of the standard solutions.
- Avoid air bubble entrapment during transfer of standard solutions.
Frequency
Once in 6 months.
QUALITY ASSURANCE DEPARTMENT
CALIBRATION REPORT OF BROOKFIELD VISCOMETER
Reference –
|
Date -
|
|||||||||
Frequency - Once in
6 Months
|
Instrument No. -
|
|||||||||
Make – Brookfield Engineering Laboratories, Inc.
|
Model – LVDV-II + Pro
|
|||||||||
Take out
the viscosity standard solutions in 500 ml beaker, adjust the temperature
to 250C & attach the required spindle, adjust required speed
& measure the viscosity, note down the constant reading.
|
||||||||||
Sr. No.
|
Viscosity
Std. in CPS
|
Actual
Viscosity of Std. in CPS
|
Observed
Viscosity in CPS
|
Tolerance
(in CPS)
|
Remark
(S / NS)
|
|||||
Minimum
|
Maximum
|
|||||||||
1
|
5.00
|
4.70
|
4.55
|
4.85
|
||||||
2
|
50.00
|
48.40
|
46.9
|
49.9
|
||||||
3
|
5000.00
|
4820.00
|
4651.83
|
4988.17
|
||||||
Conclusion : Calibration
is satisfactory (S) / not satisfactory (NS).
|
||||||||||
Next Calibration Due On :
|
||||||||||
Done
By :
Date
:
|
Checked
By :
Date
:
|
|||||||||