Calibration of Brookfield Viscometer

Learn how to verify the performance of Brookfield Viscometer using 5.00 cps, 50.00 cps and 5000.00 cps standards.

PROCEDURE

  • Ensure that all the connections of the instrument are proper.
  • Turn the viscometer ON.
  • Take out the viscosity standard solutions in 500 ml beaker, adjust the temperature to 250C. Attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.
  • Viscosity standards, speeds, spindles required & tolerance limits are as follows:
Sr.
No.
Viscosity of standard in cps
Actual viscosity of standard in cps
Spindle No.
Entry Code
   RPM
Tolerance (in cps)
Minimum
Maximum
 1.
 5.00
4.70
 ULA
00
60
4.55
4.85
 2.
 50.00
 48.40
 LV1
61
60
46.9
49.9
 3.
5000.00
4820.00
LV3
63
 10
4651.83
4988.17

Acceptance criteria

  • Viscosity should be within the tolerance limits.
  • Do not transfer the viscosity solution to the original container back after use.

Precautions

  • Take due care to avoid any contamination of the standard solutions.
  • Avoid air bubble entrapment during transfer of standard solutions.

Frequency

Once in 6 months.

QUALITY ASSURANCE DEPARTMENT
CALIBRATION REPORT OF BROOKFIELD VISCOMETER
Reference – 
Date -
Frequency - Once in 6 Months
Instrument No. -
Make – Brookfield Engineering Laboratories, Inc.
Model – LVDV-II + Pro
Take out the viscosity standard solutions in 500 ml beaker, adjust the temperature to 250C & attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.

Sr. No.
Viscosity Std. in CPS
Actual Viscosity of Std. in CPS
Observed Viscosity in CPS
Tolerance (in CPS)
Remark
(S / NS)

Minimum
Maximum

1
5.00
4.70
4.55
4.85

2
50.00
48.40
46.9
49.9

3
5000.00
4820.00
4651.83
4988.17

Conclusion : Calibration is satisfactory (S) / not satisfactory (NS).

Next Calibration Due On :

Done By :
Date :
Checked By :
Date :






Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


2 comments: Post Yours! Read Comment Policy ▼

  1. Anonymous20 April

    such a good procedure. It really help me to draft a standard calibration procedure for my organization.

    tq for sharing

    ReplyDelete
  2. How to i have know the spindle,Rpm for standard fluid in Brookfield odometer LVII+.

    ReplyDelete

Please don't spam. Comments having links would not be published.


Popular Categories

QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
Online Courses


Follow Pharmaguideline


DOCUMENTS

PHARMACEUTICAL DOCUMENTS




Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

View


adsbypg


Recent Posts