    Destruction of Microbial Waste

    Learn how to descruct the microbial waste generated from microbial analysis.

    1.0 Equipment Required

    Validated Autoclave
    SS trays
    Heat resistant Hand gloves

    2.0 Material Required

    Microbial waste
    Aluminum foil
    Chemical Indicator slip
    Nonabsorbent cotton
    Disinfectant solution.

    3.0 Procedure:

    3.1 Wear protective garments, nose mask and hand gloves prior to start the destruction of microbial waste.
    3.2 Collect and transfer all used microbiological media, cultures and contaminated glassware to washing area.
    3.3 Remove the liquid media and agar media in a waste destruction container with the help of the spatula.
    3.4 Add a disinfectant solution (except for IPA 70%) i.e. half quantity of the waste media and plug with cotton and finally wrap with aluminum foil.
    3.5 For pipettes, spatula, forceps and spreader soak in the disinfectant solution for 30 minutes and proceed for cleaning of glassware for microbiological testing.
    3.6 Leave the waste media and glassware for half an hour at room temperature to facilitate the reaction of disinfectant solution.
    3.7 For dehydrated culture media, empty out the content in a large glass container, add sufficient water to dissolve the medium and plug with non-absorbent cotton.
    3.8 Transfer the waste media and glassware to autoclave along with autoclave chemical indicator.
    3.9 Start the autoclave as per SOP for Operation of Autoclave and note the time of autoclave starting.
    3.10 Note down the time when the temperature and pressure reached 121°C and 15 psi respectively.
    3.11 Hold the cycle for 30 minutes (note down the holding time) and shut OFF an autoclave after 30 minutes of holding time.
    3.12 Open the Autoclave when the pressure becomes 0 psi, as per SOP for autoclave operation.
    3.13 Collect the autoclaved material is stainless steel tray and transfer to washbasin of washing Area.
    3.14 Drain out the media with the flow of water, rinse the glassware with water and clean the glassware as per SOP for cleaning of glassware for microbiological testing.
    3.15 Immediately run the empty autoclave load along with chemical indicator slip for 30 minutes.

    4.0 Precaution

    4.1 During loading and unloading of Autoclave always wear a safety heat resistant hand gloves.
    4.2 Do not keep the glassware rinsed with solvent.
    4.3 Do not autoclave the fresh media for routine use with destruction cycle.
    4.4 Wash the hands with soap and then rinse with 70% isopropyl alcohol after completing the destruction operations.

    5.0 Abbreviations

    SOP: Standard Operating Procedure
    QA: Quality Assurance
    QC: Quality Control
    NA: Not Applicable
    SS: Stainless steel
    WFI: Water for Injection
    Also see: Effluent Treatment Process

    1 comment: Post Yours! Read Comment Policy ▼

    1. Used, contaminated media for destruction transferred under LAF before destruction in autoclave??? plz guide me.

      ReplyDelete

    Please don't spam. Comments having links would not be published.


    Popular Categories

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses
    Follow on Whatsapp


    Follow Pharmaguideline



    DOCUMENTS

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App