- Check that all the connections of the instrument are proper.
- Ensure that printer is attached to the instrument.
- Attach sample compartment.
- Log in the software.
- Go to Set up > validation > System suitability.
- The instrument may ask for gathering the configuration, then click OK.
- System suitability set up will appear.
- Go to checks & select only Abscissa, screen report.
- Go to archive & set archive prefix as required & click OK.
Resolution
- Go to Instrument > scan > Instrument > Advanced.
- Keep Apodization ‘none’ & scan speed 0.1cm/s.
- The instrument may ask for collecting new background, then click OK.
- Go to Instrument > Validate > system Suitability & click OK.
- The instrument should display ‘System suitability test pass’.
- Scroll the report from the original path for display, click print icon & take print out.
- Abscissa spectrum of polystyrene film will be displayed.
- Go to graph format icon & select the range 1600 to 1580.
- Zoomed spectrum will appear, then collect the continuous curser & label at about 1583 cm-1 & 1589 cm-1, find out %T at these wavenumbers & find out the difference between two %T.
- The difference should be not less than 12%T.
- Similarly zoom between 2840 cm-1 to 2880 cm-1 & label at 2851 cm-1 & 2870 cm-1, find out %T at these wave numbers & find out difference.
- The difference should be not less than 18%T.
- Then adjust the scale to 4000 to 450 cm-1 & take printouts.
- Verification of wave number scale.
- Go to Process > Peak table & take its print.
- Peaks (transmission minima) should be observed at the wave numbers as follows, within the tolerance limit.
Wave number cm-1 Minimum Tolerance cm-1 Maximum Tolerance cm-1
3060.0 3058.5 3061.52849.5 2848.0 2851.0
1942.9 1941.4 1944.4
1601.2 1600.2 1602.2
1583.0 1582.0 1584.0
1154.5 1153.5 1155.5
1028.3 1027.3 1029.3
After repair calibrate the instrument before use.
1583.0 1582.0 1584.0
1154.5 1153.5 1155.5
1028.3 1027.3 1029.3
- Remove sampling compartment.
- Attach the diffuse reflectance sampling accessory.
- Collect the background with KBr.
- Go to Set up Þ validation Þ system suitability.
- Set up will appear.
- Go to checks & select only Abscissa, screen report.
- Go to archive & set archive prefix as required & click OK.
- Go to Instrument Þ validate Þ system suitability.
- Click OK.
- The instrument should display ‘System suitability test passes.
- The report will be displayed, click print icon & take print out.
Maintenance/ Repairs
Whenever the instrument is out of order, then put the tag as ‘Out of Calibration’ & get it repaired.After repair calibrate the instrument before use.
Frequency
Once in three months.Attachments
Enclose following attachments with the calibration report.- System suitability check report for slide holder instrument configuration.
- Abscissa spectrum of polystyrene film with identified peaks’ table for wave number calibration.
- Abscissa spectrum of polystyrene film (showing %T at 2851cm-1, 2870cm-1, 1583 cm-1, 1589 cm-1).
- System suitability check report for diffuse reflectance instrument configuration.
- Note: Words in italics stand for instrument tabs.
CALIBRATION REPORT OF FT-IR SPECTROPHOTOMETER
|
||||||||
Reference –
|
Date -
|
|||||||
Frequency - Once in three Months.
|
Instrument No. -
|
|||||||
Make –
|
Model –
|
|||||||
1.0
|
Resolution:
|
|||||||
Difference Between % T at 2851 cm-1 and 2870 cm-1
|
Difference Between % T at 1583 cm-1 and 1589 cm-1
|
|||||||
Limit
|
NLT 18%
|
NLT 12%
|
||||||
Actual
|
||||||||
2.0
|
Verification of Wave Number Scale -
|
|||||||
Observed Wave Number cm-1
|
Minimum Tolerance
cm-1
|
Maximum Tolerance
cm-1
|
||||||
3058.5
|
3061.5
|
|||||||
2848.0
|
2851.0
|
|||||||
1941.4
|
1944.4
|
|||||||
1600.2
|
1602.2
|
|||||||
1582.0
|
1584.0
|
|||||||
1153.5
|
1155.5
|
|||||||
1027.3
|
1029.3
|
|||||||
3.0
|
System suitability after attachment of diffuse reflectance sampling accessory : Complies / Does Not Comply.
|
|||||||
4.0
|
Conclusion : Calibration is satisfactory / Not satisfactory.
|
|||||||
5.0
|
Next Calibration Due On :
|
|||||||
Done By :
Date :
|
Checked By :
Date :
|
|||||||