Calibration of FTIR Spectrophotometer

Know the procedure to calibrate the FTIR Spectrophotometer used in identification of Pharmaceutical products.
  • Check that all the connections of the instrument are proper. 
  • Ensure that printer is attached to the instrument. 
  • Attach sample compartment. 
  • Log in the software. 
  • Go to Set up > validation > System suitability. 
  • The instrument may ask for gathering the configuration, then click OK. 
  • System suitability set up will appear. 
  • Go to checks & select only Abscissa, screen report. 
  • Go to archive & set archive prefix as required & click OK. 

Resolution

  • Go to Instrument > scan > Instrument > Advanced.
  • Keep Apodization ‘none’ & scan speed 0.1cm/s.
  • The instrument may ask for collecting new background, then click OK.
  • Go to Instrument > Validate > system Suitability & click OK.
  • The instrument should display ‘System suitability test pass’.
  • Scroll the report from the original path for display, click print icon & take print out.
  • Abscissa spectrum of polystyrene film will be displayed.
  • Go to graph format icon & select the range 1600 to 1580.
  • Zoomed spectrum will appear, then collect the continuous curser & label at about 1583 cm-1 & 1589 cm-1, find out %T at these wavenumbers & find out the difference between two %T.
  • The difference should be not less than 12%T.
  • Similarly zoom between 2840 cm-1 to 2880 cm-1 & label at 2851 cm-1 & 2870 cm-1, find out %T at these wave numbers & find out difference.
  • The difference should be not less than 18%T.
  • Then adjust the scale to 4000 to 450 cm-1 & take printouts.
  • Verification of wave number scale.
  • Go to Process > Peak table & take its print.
  • Peaks (transmission minima) should be observed at the wave numbers as follows, within the tolerance limit.
Wave number  cm-1     Minimum Tolerance cm-1     Maximum Tolerance cm-1
3060.0                            3058.5                                     3061.5
2849.5                            2848.0                                     2851.0
1942.9                            1941.4                                     1944.4

1601.2                            1600.2                                     1602.2
1583.0                            1582.0                                     1584.0
1154.5                            1153.5                                     1155.5
1028.3                            1027.3                                     1029.3

  • Remove sampling compartment. 
  • Attach the diffuse reflectance sampling accessory. 
  • Collect the background with KBr. 
  • Go to Set up Þ validation Þ system suitability. 
  • Set up will appear. 
  • Go to checks & select only Abscissa, screen report. 
  • Go to archive & set archive prefix as required & click OK. 
  • Go to Instrument Þ validate Þ system suitability
  • Click OK. 
  • The instrument should display ‘System suitability test passes. 
  • The report will be displayed, click print icon & take print out.

Maintenance/ Repairs

Whenever the instrument is out of order, then put the tag as ‘Out of Calibration’ & get it repaired.
After repair calibrate the instrument before use.

Frequency

Once in three months.

Attachments

Enclose following attachments with the calibration report.
  • System suitability check report for slide holder instrument configuration. 
  • Abscissa spectrum of polystyrene film with identified peaks’ table for wave number calibration. 
  • Abscissa spectrum of polystyrene film (showing %T at 2851cm-1, 2870cm-1, 1583 cm-1, 1589 cm-1). 
  • System suitability check report for diffuse reflectance instrument configuration. 
  • Note: Words in italics stand for instrument tabs.
CALIBRATION REPORT OF FT-IR SPECTROPHOTOMETER
Reference –
Date -
Frequency - Once in three Months.
Instrument No. -
Make – 

  Model – 

1.0
Resolution:
 
Difference Between % T at 2851 cm-1 and 2870 cm-1
Difference Between % T at 1583 cm-1 and 1589 cm-1
Limit
NLT 18%
NLT 12%
Actual
2.0
Verification of Wave Number Scale -
Observed Wave Number cm-1
Minimum Tolerance
cm-1
Maximum Tolerance
cm-1
3058.5
3061.5
2848.0
2851.0
1941.4
1944.4
1600.2
1602.2
1582.0
1584.0
1153.5
1155.5
1027.3
1029.3
3.0
System suitability after attachment of diffuse reflectance sampling accessory : Complies / Does Not Comply.
4.0
Conclusion : Calibration is satisfactory / Not satisfactory.
5.0
Next Calibration Due On :
Done  By :
Date :
Checked By :
Date :













Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


2 comments: Post Yours! Read Comment Policy ▼

  1. Anonymous05 May

    This is exactly what I've been looking for.

    Thanks!

    ReplyDelete
  2. In the SOP of my labs tolerance limits for 1942, 2849 and 3060 wavelength is just ± 1.0, however in this example you use ±1.5, which is the source of that difference. Because my equipment has -1.13 at 3060.0. Thanks in advance

    ReplyDelete

Please don't spam. Comments having links would not be published.


Popular Categories

QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
Online Courses


Follow Pharmaguideline


DOCUMENTS

PHARMACEUTICAL DOCUMENTS




Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

View


adsbypg


Recent Posts