1.0 PURPOSE
To lay down the procedure for the operation of Count Down Timer.2.0 SCOPE
It is applicable to Microbiology laboratory.3.0 RESPONSIBILITY
Microbiologist4.0 ACCOUNTABILITY
Head of Department5.0 PROCEDURE
5.1 Checks before starting of instrument
5.1.1 Check the calibration status before using the instrument.5.1.2 Check the cleaning status of the instrument.
5.2 Operation
5.2.1 Press button “ Hour/Min” to set Hour /minute/Function.5.2.2 Press button “Min/Sec” to set Minute/Second function.
5.2.3 Press button “Start/Stop” to start counting down.
5.2.4 To temporarily stop timing cycle, press button “Start/Stop”.
5.2.5 Press button “ Hour/Min” and “Minute/Sec” at the same time to clear the digit and reset the Hour/Min and Minute/Sec.
5.3 The minimum time set is for one minute.
5.4 When set time is up, the alarm will sound.
5.5 Press any button to stop alarm sound, otherwise the sound will automatically stop within 30 sec.
6.0 ABBREVIATIONS
6.1 SOP – Standard operating procedure6.2 Min – Minute
6.3 Sec – Second
Get ready to use editable documents in MS-Word FormatView List
No comments:
Post a Comment
Please don't spam. Comments having links would not be published.