1.0 Purpose
To lay down the procedure to monitor compressed air/gases for microbiological, non-viable particulate, oil and moisture content.2.0 Scope
Applicable for monitoring of compressed air and gases (Nitrogen, Oxygen and Carbon Dioxide) in different production facilities for microbiological, non-viable particulate, oil and moisture content.3.0 Responsibility
Trained QC – Microbiology Officer is responsible for performing tests as per SOP.4.0 Accountability
Head of Department5.0 Procedure
5.1 Microbiological Monitoring
Note: Microbiological monitoring is performed by sterility testing for user points in the aseptic zone (filling and filtration area) and by Total Aerobic Microbial Counts (using M Air T CG Sampler) for other user points.5.1.1 Test for Sterility
5.1.1.1 Prepare the Soybean Casein Digest medium as per SOP for media preparation.5.1.1.2 Dispense 100 ml of Soybean Casein Digest Medium in cleaned gas sampling assembly. Close the inlet and outlet tubes with air 0.2µ vent filters.
5.1.1.3 Sterilize the assembly as per SOP.
5.1.1.4 Pre - Incubate the assembly with media at 30 – 35°C for 48 hours before use.
5.1.1.5 Bring the sample assembly to sampling location.
5.1.1.6 Disinfect the outer surface of sampling point by the spraying of 70% v/v isopropyl alcohol or recommended disinfected solution
5.1.1.7 Flush the gas sample for 1 min.
5.1.1.8 Remove the vent filter from the inlet tubing and attach to filter outlet connected to the sampling point.
5.1.1.9 Open the air/gas sample valve slowly and take the sample for 10 min.
5.1.1.10 After collecting the sample replace the vent filter of inlet tubing and label the assembly with appropriate details viz. Sample Identification, Date and Initials of the analyst.
5.1.1.11 Bring the sample to the laboratory and incubate assembly at 22.5 ± 2.5 oC for 7 days followed by incubation at 32.5 ± 2.5 oC for 7 days. In case of Nitrogen Samples, incubate the samples in anaerobic conditions.
5.1.1.12 After completion the incubation period record the results.
5.1.1.13 Frequency: Once in One month
5.1.1.14 Acceptance Criteria: No Growth Should be Observed
5.1.2 Test for Total Aerobic Microbial Counts
5.1.2.1 Transfer the pre incubated Soybean Casein Digest Agar cassette, M Air T CG Sampler and accessories to the area to be monitored in a closed container to avoid any contamination.5.1.2.2 Write the sampling location ID and date of sampling on the base of cassette.
5.1.2.3 Connect the sampling point to M Air T CG Sampler to and perform sampling as per SOP for 1000 litres.
5.1.2.4 After completion of air sampling close the cassette with lid.
5.1.2.5 Collect all exposed cassette in a closed container and transfer the plates to incubators for incubation.
5.1.2.6 Incubate the exposed cassettes along with negative control in inverted position at 30-35oC for 5 Days. In case of Nitrogen Samples, incubate the samples in anaerobic conditions.
5.1.2.7 After incubation observe and count the number of colonies and record the results.
5.1.2.8 Frequency: Once in One month
5.1.2.9 Acceptance Criteria: Shall Comply for Room Classification Limits as follows:
Grade
|
Action Limit
CFU/m3
|
Grade A
|
1
|
Grade B
|
10
|
Grade C
|
100
|
Grade D
|
200
|
Other Grades (Grade E, O, etc)
|
For Information
|
5.2 Non – Viable Particulate Count
5.2.1 Take the air-borne particle counter (LASAIR) and accessories to the respective sampling rooms.5.2.2 Connect the user point to a stainless steel box (with inlet and outlet) to diffuse the high pressure of air/gas to appropriate pressure to enable counting. Slowly open the user point and allow some air/gas to flush the box.
5.2.3 Place the sampling probe of the particle counter at the outlet of the SS box and sample for 1-cubic feet of air or volume.
5.2.4 Take the print of the particle counter and record results.
5.2.5 Frequency: Once in One month
5.2.6 Acceptance Criteria: Shall Comply for Room Classification Limits as follows:
Grade
|
Particles / ft3
|
|
0.5 µ
|
5.0 µ
|
|
Grade A
|
100
|
1
|
Grade B
|
100
|
1
|
Grade C
|
10000
|
57
|
Grade D
|
100000
|
571
|
Other Grades (Grade E, O, etc)
|
For Information
|
5.3 Test for Oil and Moisture Content
Note: These tests are applicable for Compressed air points only.5.3.1 This test shall be performed by the external testing laboratory as per protocol or by using gas detector tubes (Gastec).
5.3.2 Frequency
Sample Point from Compressed Air Generation System: Once a MonthProcess User Points: One point monthly on rotation
5.3.3 Acceptance Criteria
Oil: £ 0.5 mg/m3Moisture: £ 0.5 g/m3
6.0 ABBREVIATIONS
6.1 SOP - Standard Operating Procedure6.2 % - Percent
6.3 °C - Degree centigrade
6.4 CFU - Colony Forming Unit
6.5 ft3 - Cubic Feet
6.6 m3 - Cubic Meter
6.7 g - Gram
6.8 mg - Milligram
Get ready to use editable documents in MS-Word FormatView List
No comments:
Post a Comment
Please don't spam. Comments having links would not be published.