1.0 Objective
To check the performance of the equipment for reliable and accurate results.2.0 Scope
This procedure is applicable to check the performance of the atomic absorption spectrophotometer installed in the quality control department.3.0 Responsibility
3.1 Doing: Technical Assistant3.2 Checking: Executive /Manager
4.0 Accountability
Head of the Department5.0 Procedure
Frequency: Once in a month5.1 Operate the instrument as per respective standard operating procedure.
5.2 Use Cu standard for radiations.
5.3 Prepare 0.5 mcg/ml, 1.0 mcg/ml, 1.5 mcg/ml, 2.0 mcg/ml and 2.5 mcg/ml in water.
5.4 Aspirate the above solution and measure the absorbance.
5.5 Calculate the correlation coefficient by the instrument/computer.
5.6 Correlation coefficient should not less than 0.9900 %.
5.7 The performance of the instrument is satisfactory if the obtained correlation coefficient is within the given limit otherwise follow the S.O.P.
6.0 ABBREVIATIONS
6.1 % = Percentage6.2 mcg= microgram
6.3 ml= millilitre
6.4 cu= copper
ANNEXURE-I
PERFORMANCE CHECK REPORT OF ATOMIC ABSORPTION
PERFORMANCE CHECK DATE
|
DATE OF LAST PERFORMANCE CHECK DONE
|
NEXT DUE FOR
PERFORMANCE CHECKE
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
Test % w/v
|
Absorption
|
Mean
| ||
1
|
2
|
3
| ||
0.5 mcg/ml
| ||||
1.0 mcg/ml
| ||||
1.5 mcg/ml
| ||||
2.0 mcg/ml
| ||||
2.5 mcg/ml
|
Correlation coefficient : _________ (Limit : Not less than 0.9900 %)