1.0 OBJECTIVE
To describe the procedure for the calibration of the hot air oven2.0 SCOPE
This SOP is applicable for the calibration of hot air oven.3.0 RESPONSIBILITY
Officer /Executive - Quality Control4.0 ACCOUNTABILITY
Manager - Quality Control.5.0 PROCEDURE
5.1 For the calibration of hot air oven use a standard thermometer ranging up to 300º C.5.2 Start the calibration procedure after 1 hour of staring the oven.
5.3 Set the oven at desired temperature.
5.4 Put the standard thermometer for 30 minutes on the upper shelf of the oven and close the door of the oven.
5.5 After 30 minutes open the door of the oven and read the temperature of standard thermometer randomly and match the observed temperature of the thermometer.
5.6 Repeat the above procedure by putting the thermometer in lower shelf for 30 minutes.
5.7 Record the observed temperature in calibration record of hot air oven as per Annexure no. I
5.8 The observed temperature of the thermometer in both shelves should be ± 2.0º C tolerance limit to the set temperature value.
5.9 Frequency: Once in a month.
Related: SOP for Validation of Oven for Depyrogenation
6.0 ABBREVIATIONS
6.1 SOP - Standard Operating Procedure
ANNEXURE- I
CALIBRATION RECORD OF HOT AIR OVEN
Inst. ID No.
|
:
|
Ref. SOP No.
|
:
|
||
Make
|
:
|
Model No.
|
:
|
Date of Calibration :
Next Due Date :
Location :
Frequency: Monthly :
Location
|
Set Temperature
|
Observed Temperature
|
Tolerance
|
Observed By
|
Upper shelf
|
±2.0º C
|
|||
Lower shelf
|
±2.0º C
|
Remark: The calibration procedure of instrument are within tolerance limit /out of tolerance limit and gives satisfactory / not satisfactory results.
Analyst Checked by
Date Date
Get ready to use editable documents in MS-Word FormatView List
No comments:
Post a Comment
Please don't spam. Comments having links would not be published.