SOP for Calibration of Glassware

Standard operating procedure to calibrate the volumetric glassware used in Pharmaceutical analysis.

1.0 OBJECTIVE

To describe the procedure for calibration of glassware.

2.0 SCOPE

This SOP is applicable for calibration of glassware.

3.0 RESPONSIBILITY

Officer/ Executive - Quality Control

4.0 ACCOUNTABILITY

Manager – Quality Control

5.0 PROCEDURE

5.1 Volumetric glassware’s are calibrated at 23° to 27°C.
5.2 Pharmacopoeial assay and other tests involving volumetric measurements require the use of accuracy-calibrated glassware.
5.3 Clean the glassware for calibration to remove any grease, dirt etc. by keeping it immersed in chromic acid for 30 minutes.
5.4 Rinse the glassware under running tap water then rinse it twice with de-ionized water and dry it at 60-65°C for 60 minutes. Ensure that glassware is completely dry.
5.5 Remove the glassware from the oven and keep it in desiccators, ensure that the ambient temperature is 25°C ± 2°C.
5.6 Arrange a water bath consisting of plastic beaker contain water at 25°C ± 2°C.
5.7 Maintain the temperature of water bath at 25°C ± 0.5°C throughout the calibration.
5.8 Place the apparatus to be calibrated on the balance and note the weight (A), remove the apparatus and transfer the distilled water at 25°C so that the meniscus of liquid in the apparatus on the mark level. Weight the apparatus and note down the weight (B).
5.9 Calculate the volume of glass ware using following formula i.e. (B – A)/ 0.99602
5.10 Take the volumetric flask for calibration (10 to 1000 ml)
5.11 Take pipette for calibration (1 to 50 ml)
5.12  Volumetric Flask:
         Nominal Capacity, (ml)           10         25         50         100       200       250       500         1000
         Tolerance         ±    (ml)          0.02      0.03      0.04      0.06      0.08      0.10      0.15         0.20
         (e.g., for 10 ml volumetric flask)
         Wt taken of volumetric flask = W gm
         Wt of volumetric flask + water (upto the mark level)= W1 gm
         Wt of water  = W1 – W gm
5.11  One Mark Pipette (bulb pipette)
         Nominal Capacity,  (ml)            1          2           5          10         20         25         50
         Tolerance        ±     (ml)          0.01     0.01     0.02      0.02      0.03      0.03      0.04
5.12  Graduated Pipette
         Nominal Capacity,  (ml)             1           2            5          10         25
         Sub division            (ml)           0.01      0.02      0.05       0.1        0.1                       
         Tolerance        ±     (ml)           0.01      0.01      0.02      0.02      0.03
5.13  Burette
         Nominal Capacity,  (ml)         25         50
         Sub division            (ml)        0.05      0.1                       
         Tolerance        ±     (ml)        0.03      0.05

6.0 ABBREVIATIONS

6.1 SOP - Standard Operating Procedure
6.2 Ml - Milliliter
Get ready to use editable documents in MS-Word FormatView List





Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


No comments:

Post a Comment

Please don't spam. Comments having links would not be published.


Popular Categories

QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
Online Courses


Follow Pharmaguideline


DOCUMENTS

PHARMACEUTICAL DOCUMENTS




Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

View


adsbypg


Recent Posts