Check HPLC chromatography (Pump) for the following:
I. Checkpoint: Leakage test (By Pressure Drop)
II. Flow rate calibration
2. Place inlet tubing of the Pump into the Water HPLC grade through a suction filter.
I. Checkpoint: Leakage test (By Pressure Drop)
II. Flow rate calibration
CHECKPOINT: LEAKAGE TEST (BY PRESSURE DROP)
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.2. Place inlet tubing of the Pump into the Water HPLC grade through a suction filter.
3. Allow the mobile phase to flow for about 5 min.
4. Block Pump outlet with the block screw.
5. The pressure rises and on crossing the 300 bar, “ERROR P-MAX” appears on the display window. Note the time. Press “CE” key and observe the pressure drop for 5 min.
6. After 5 min., record the pressure in the Calibration Log.
7. Make entry of the HOLC column usage in the Column Usage Log Register.
8. Make entry of the user into the Instrument Usage Log Register.
9. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the HPLC chromatography Calibration Status.
4. Block Pump outlet with the block screw.
6. After 5 min., record the pressure in the Calibration Log.
7. Make entry of the HOLC column usage in the Column Usage Log Register.
8. Make entry of the user into the Instrument Usage Log Register.
9. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the HPLC chromatography Calibration Status.
10. In case of non-compliance, follow the Maintenance Program.
Related: Principle and Working of HPLC Chromatography
2. Ensure that, the Pump is passing the “Leakage Test (By Pressure Drop)”.
3. Keep the Drain tube in such a way that the mobile phase (Water) drops falls into 10 ml clean, dry volumetric flask without touching the walls of the flask and start immediately the stopwatch when the first drop falls into the flask.
4. Wait till the collected mobile phase reaches 10 ml mark of the volumetric flask and Stop the stopwatch.
5. Record the time required to collect the 10 ml mobile phase in the calibration log.
6. Repeat the procedure for 1.0 ml, 1.5 ml and 2.0 ml/ min. flow rates.
7. Repeat step 3 to 6 but using methanol HPLC grade as mobile phase instead of water.
8. Compare the results for its compliance with limits given in the Calibration Log and put the remark regarding the Calibration Status.
9. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program.
2. Pipette out 1.0 ml of Benzene and Toluene into the clean and dry 50 ml volumetric flask.
3. Make up the volume to 50 ml with Methanol and mix well.
Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 µl
2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject 10 µl in triplicate and record the chromatograms.
4. Repeat the injection of the above solution by injecting 15, 20, 25, 30 µl in triplicate.
5. Take the print out of the chromatograms and attach to the Calibration Log.
6. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
7. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
8. Plot the curve for the area corresponding to Benzene to Toluene peaks v/s injection volume, Find out the RSD (reproducibility) and record in the Calibration Log.
9. Find out the Correlation coefficient “r2” for each peak at five levels and record in the Calibration Log.
10. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
11. Prepare Calibration Status Label and display on the instrument at the designated place.
12. In case of non-compliance, follow the Maintenance Program.
2. On the Detector’s display window, some values at the functions “l(nm)”, “abs(AU)”, “range(AUFS)” and “lamp” appears.
3. On the display, previously set value blinks at “l(nm)” function, enter the wavelength to 254 nm by pressing numeric keys.
4. Press “Func Back” key and select “lamp” functions, enter 1 to select D2 lamp.
5. Further press “Func Back” key till “REF EN ” appears.
6. Record the Reference Energy of the D2 Lamp at 254 nm in the Calibration Log.
7 Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
8. In case of non-compliance, follow the Maintenance Program.
2. Pipette out ml of Benzene and Toluene as per the following table into the clean, dry 50 ml volumetric flask separately.
3. Make up the volume to 50 ml with Methanol and mix well.
HPLC Column: ODS C18, (25 cm x 4.6 mm ID, 5 m)
Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 µl
2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject each of the Solutions in triplicate and record the chromatograms.
4. Take the print out of the chromatograms and attach to the Calibration Log.
5. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
6. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
7. Calculate the Area Ratio of Benzene to Toluene, Find out the Mean value of the ratios and record in the Calibration Log.
8. Find out the Correlation coefficient “r2” from the Mean area ratio values of the three levels.and record in the Calibration Log.
8. Plot the curve for area ratio corresponding to Benzene to Toluene peaks v/s concentration, Find out the RSD (reproducibility) and record in the Calibration Log.
9. Compare the result for its compliance against limit given in the Calibration Log and put the remark regarding the Calibration Status.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program.
Instrument Code No.:
Related: Principle and Working of HPLC Chromatography
FLOW RATE CALIBRATION
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.2. Ensure that, the Pump is passing the “Leakage Test (By Pressure Drop)”.
3. Keep the Drain tube in such a way that the mobile phase (Water) drops falls into 10 ml clean, dry volumetric flask without touching the walls of the flask and start immediately the stopwatch when the first drop falls into the flask.
4. Wait till the collected mobile phase reaches 10 ml mark of the volumetric flask and Stop the stopwatch.
5. Record the time required to collect the 10 ml mobile phase in the calibration log.
6. Repeat the procedure for 1.0 ml, 1.5 ml and 2.0 ml/ min. flow rates.
7. Repeat step 3 to 6 but using methanol HPLC grade as mobile phase instead of water.
8. Compare the results for its compliance with limits given in the Calibration Log and put the remark regarding the Calibration Status.
9. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program.
REPRODUCIBILITY AND LINEARITY OF INJECTION VOLUME
Solution Mixer Preparation
1. Take clean and dry 50 ml volumetric flask.2. Pipette out 1.0 ml of Benzene and Toluene into the clean and dry 50 ml volumetric flask.
3. Make up the volume to 50 ml with Methanol and mix well.
Chromatographic Condition
HPLC Column: ODS C18, (25 cm x 4.6 mm ID, 5 µm)Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 µl
Calibration
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject 10 µl in triplicate and record the chromatograms.
4. Repeat the injection of the above solution by injecting 15, 20, 25, 30 µl in triplicate.
5. Take the print out of the chromatograms and attach to the Calibration Log.
6. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
7. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
8. Plot the curve for the area corresponding to Benzene to Toluene peaks v/s injection volume, Find out the RSD (reproducibility) and record in the Calibration Log.
9. Find out the Correlation coefficient “r2” for each peak at five levels and record in the Calibration Log.
10. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
11. Prepare Calibration Status Label and display on the instrument at the designated place.
12. In case of non-compliance, follow the Maintenance Program.
CHECKPOINT: D2 LAMP ENERGY CHECK (DETECTOR)
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.2. On the Detector’s display window, some values at the functions “l(nm)”, “abs(AU)”, “range(AUFS)” and “lamp” appears.
3. On the display, previously set value blinks at “l(nm)” function, enter the wavelength to 254 nm by pressing numeric keys.
4. Press “Func Back” key and select “lamp” functions, enter 1 to select D2 lamp.
5. Further press “Func Back” key till “REF EN ” appears.
6. Record the Reference Energy of the D2 Lamp at 254 nm in the Calibration Log.
7 Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
8. In case of non-compliance, follow the Maintenance Program.
III. LINEARITY OF DETECTOR RESPONSE :
Solution Mixer Preparation:
1. Take three clean and dry 50 ml volumetric flasks.2. Pipette out ml of Benzene and Toluene as per the following table into the clean, dry 50 ml volumetric flask separately.
3. Make up the volume to 50 ml with Methanol and mix well.
Solution (Level)
|
ml of Benzene to be taken
|
Ml of Toluene to be taken
|
To be Diluted with Methanol to
|
1
|
0.5
|
1.0
|
50 ml
|
2
|
1.0
|
1.0
|
50 ml
|
3
|
1.5
|
1.0
|
50 ml
|
Chromatographic Condition
1. Set each of the equipment to the following parameters as per the Equipment’s Parameter Setup Procedure.HPLC Column: ODS C18, (25 cm x 4.6 mm ID, 5 m)
Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 µl
Calibration
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject each of the Solutions in triplicate and record the chromatograms.
4. Take the print out of the chromatograms and attach to the Calibration Log.
5. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
6. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
7. Calculate the Area Ratio of Benzene to Toluene, Find out the Mean value of the ratios and record in the Calibration Log.
8. Find out the Correlation coefficient “r2” from the Mean area ratio values of the three levels.and record in the Calibration Log.
8. Plot the curve for area ratio corresponding to Benzene to Toluene peaks v/s concentration, Find out the RSD (reproducibility) and record in the Calibration Log.
9. Compare the result for its compliance against limit given in the Calibration Log and put the remark regarding the Calibration Status.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program.
CALIBRATION OF HPLC ( Liquid Chromatography ) PUMP
Model No. :
|
Make :
|
Instrument Code No. :
|
A. Checkpoint : Leakage Check Test (Pressure Drop)
| |||||||||||
P.Max
to be set
|
P.Max Set
|
Start Time
|
Stop Time
|
Total Time
|
Pressure Observed
|
Remark
| |||||
300 bar
| |||||||||||
B. Flow Rate calibration :
| |||||||||||
Flow Rate
(ml / min.) | Time required to collect 10 ml mobile phase | Limit | Remarks | ||||||||
Theoretical
(in sec.)
| Actual with water | Actual with Methanol | |||||||||
0.5 ml | 1200 | 1194 - 1206 | |||||||||
1.0 ml | 600 | 594 – 606 | |||||||||
1.5 ml | 450 | 443 – 457 | |||||||||
2.0 ml | 300 | 294 - 306 | |||||||||
Calibration Status : Satisfactory / Not Satisfactory
Next Calibration Due :
Calibrated By: Checked By :
Date: Date:
CALIBRATION OF INJECTOR
Model: Make:Instrument Code No.:
a. Preparation of solvent mixture :
Taken _____ ml (1.0 ml) of the ________ (Benzene) and ________ (Toluene) in to a _____ ml (50 ml) clean, dry volumetric flask, make up the volume with methanol, mixed well.b. Chromatographic Condition :
Parameters | Test Condition | Applied Condition |
Mobile Phase |
Methanol: Water (70: 30)
| |
HPLC Column | ODS C18, (25 cm x 4.6 mm ID, 5 µm) | |
Flow Rate | 1.0 ml / min. | |
Wavelength | 254 nm |
Injection volume | Injection -1 | Injection -2 | Injection -3 | Mean | % RSD :(Not more than 2.0%) | ||||
Benzene | Toluene | Benzene | Toluene | Benzene | Toluene | ||||
10 ml | RT | ||||||||
Area | |||||||||
15 ml | RT | ||||||||
Area | |||||||||
20 ml | RT | ||||||||
Area | |||||||||
25 ml | RT | ||||||||
Area | |||||||||
30 ml | RT | ||||||||
Area |
Coefficient of co-relations : r2 : _____________ (Limit : NLT : 0.999)
Calibration Status : Satisfactory / Not Satisfactory
Next Calibration Due :
Calibrated By: Checked By :
Date: Date :
CALIBRATION OF DETECTOR
Model :
|
Make :
|
Instrument Code No. :
|
D2 LAMP ENERGY CHECK :
| |||
Wavelength to be set
|
Wavelength Set
|
Reference Energy Observed
|
Limit
|
254 nm
|
Not less than 200
|
LINEARITY OF DETECTOR RESPONSE :
| ||||||
a. Solution Preparation : | ||||||
Solution
|
ml of Benzene
|
ml of Toluene
|
in Methanol
| |||
(Level)
|
To be taken
|
Taken
|
To be taken
|
Taken
|
To be diluted
|
Diluted to
|
1
|
0.5 ml
|
1.0 ml
|
50 ml
| |||
2
|
1.0 ml
|
1.0 ml
|
50 ml
| |||
3
|
1.5 ml
|
1.0 ml
|
50 ml
|
Mobile Phase Preparation : Taken ______ ml of ___________ (Methanol HPLC grade) in to a _______ ml clean, dry volumetric flask, added ____________ ml of ___________(Water HPLC grade), mixed well, allowed to cool to room temperature. Filtered through 0.45/0.22 m membrane filter, degassed for _____ min. by vacuum / sonication.
Parameters | Test Condition | Applied Condition |
Mobile Phase | Methanol: Water ( 70: 30) | |
HPLC Column | ODS C18, (25 cm x 4.6 mm ID, 5 µm) | |
Flow Rate | 1.0 ml / min. | |
Wavelength | 254 nm | |
Range | 1.0 AUFS | |
Injection Volume | 20 µl |
Solution(Level) | Benzene Area | Toluene Area | Area Ratio
(Benzene/Toluene)
| ||
RT | Area | RT | Area | ||
1 | |||||
Mean : | |||||
% RSD :
| --- | --- | |||
2 | |||||
Mean : | |||||
% RSD :
| |||||
3 | |||||
Mean : | |||||
% RSD :
|
Linearity of Ratio : Correlation Co-efficient r2 :
|
Limit : % RSD of the Retention Time : Not more than 2.0%
Linearity of Ratio : Correlation Co-efficient r2 : Not less than 0.999
|
Calibration Status : Satisfactory / Not Satisfactory
Next Calibration Due :
Calibrated By : Checked By :
many many thanks
ReplyDeleteWhat is the difference between calibration and performance check?
ReplyDeleteSome time caffeine is using for the calibration why caffeine using for calibration?
ReplyDeleteCaffeine is very stable easyly avilable in market compound have two high wave lenths one low wave length and low degradable product
DeleteVery good Very useful for HPLC users. Expect such helpful information in future. Thank you.
ReplyDeletewhat is the importance of carryover test in calibration? why we have to do?
ReplyDeleteCarryover test proves that the area quantity of the second injection is not influenced by the first injection. This helps in showing the accuracy and integrity of each injection area to be true.
DeleteWhy we use benzen with toluene. Why we does not use only toluene. What is the reason. Plz tell
ReplyDeleteCan you please tell me frequently asked hplc based questions in interview
ReplyDelete