CALIBRATION OF TOP PAN BALANCE
Calibration procedure
Internal calibration
It shall be performed as per the manufacturer’s instructions.Drift check
For the calculation of the drift 10 mg weight shall be used. Note down the 10 measurements of the 10 mg weight in the performance check log. (Attachment no. 01) Variation in the observed weight from the mean value does not exceed ± 0.2 mg.
The 10 mg weight shall meet the performance check criteria of the mass value i.e. 0.1% of actual mass value. E.g. For all the 10 measurements of 10 mg weight, variation in the weighing cannot exceed 0.01mg.
Performance check
After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The measurement shall be within the 0.1% of the actual mass value of the individual weight as given in the performance check log. (Attachment no. 02)
Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement Uncertainity = Standard Deviation X 2
Actual Mass Value
Measurement of uncertainty shall be Not More Than 0.10.
Measurement uncertainty check
The measurement of uncertainty shall be carried out by using 10 mg weight.Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement Uncertainity = Standard Deviation X 2
Actual Mass Value
Measurement of uncertainty shall be Not More Than 0.10.
Calibration frequency
It shall be performed daily, after any maintenance, after the relocation and after the power failure of the balance.
Each manufacturer supplies the maximum loading condition of the balance. The calibration of the balance shall be designed in such a way that the performance check weights cover the entire loading range of the balance. Lower and higher load limits shall be checked for the performance check. Before calibration make sure that the level bubble is in the center of the indicator and if internal calibration facility is available in the balance then it shall also be performed daily before performance check of the balance.
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.
Attachment-1: Performance Check Log of Microbalance
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.
Attachment-1: Performance Check Log of Microbalance
Instrument:_____________________ Code No :______________
Make :_____________________ Model :_______________
Date
|
Level indicator check
|
1 mg
|
2 mg
|
5 mg
|
10 mg
|
20 mg
|
Weight box no. and valid upto
|
Done by
|
Checked
by
|
Remark
| |
Std. Wt. Cal No. Þ
| |||||||||||
Limit
Þ
|
Bubble should be in Center
|
+ 0.1% of
Std. Weight
( to )
|
+ 0.1% of
Std. Weight
( to )
|
+ 0.1% of
Std. Weight
( to )
|
+ 0.1% of Std. Weight
( to )
|
+ 0.1% of Std. Weight
( to )
| |||||
Attachment-2: Auto Performance Check Log of Microbalance
Instrument:_____________________ Code No :______________
Make :_____________________ Model :_______________
Check level indicator (Bubble should be in center) : _________________________________
(Paste Performance check chart here)
Calibrated by : _______________ Date : ____________________
Checked by : _______________ Date : ____________________
Attachment-3: Measurement Uncertainity Check of Microbalance
Instrument:_____________________ Code No :______________
Make :_____________________ Model :_______________
Date
|
Std. Wt.
|
Std. Wt. Cal. No.
|
Observation
|
of Uncertainty
Less than or equal to 0.001
|
Weight box no. and valid upto
|
Done by
|
Checked by
|
Remark
| ||||
10 mg
| ||||||||||||
Calculation
Measurement = Standard Deviation X 2
Measurement = Standard Deviation X 2
uncertainty Actual Mass Value
Can we take weight for analysis in-between of monthly calibration?
ReplyDeleteYou do not take weight in between your balance during calibration like monthly calibration.
DeleteFirstly you complete the calibration of balance based upon evolution of calibration result you can use further for routine use.
What do u mean by actual mass value in measurement of uncertainty?
ReplyDeleteIt means actual weight.
DeleteCertified value of the standard
DeleteExactly, it is certified weight of the standard being used.
Deletehello sir,
ReplyDeleteverification of balance is necessary after every power failuar.
Reply plz. From syed sabir
Yes and if you change the location or move the balance you need to do calibration
DeleteVerification of balance is necessary after power failure and movement of balance from one to another place.
ReplyDeleteWhats the mean of drift.
ReplyDeleteDrift is error in measurements.
DeleteIs eccentricity test not required?
ReplyDeleteEccentricity test is required in balance calibration. And weight value of accentricity test in 1/3 of your maximum balance capacity
Delete