1.0 PURPOSE
To lay down the procedure to perform the Operation of Bio-Safety cabinet.2.0 SCOPE
It is applicable to Quality control department.3.0 RESPONSIBILITY
Microbiologist4.0 ACCOUNTABILITY
Head of Department5.0 PROCEDURE
5.1 Operation of Bio-Safety cabinet
5.1.1 Switch on the main supply.5.1.2 Ensure equipment is in a clean state.
5.1.3 Mop the bench with 70 % isopropyl alcohol/ disinfectant.
5.1.4 Start blowers 30 minutes prior to the operation.
5.1.5 Switch “ON” the Ultra Violet (U.V) light for 1 hour prior to use or when required.
5.1.6 Ensure that when Ultra Violet (U.V) light is in “ON” condition, the normal light should be switched “OFF”.
5.1.7 Record the differential pressure across the High efficiency Particulate Air (HEPA) filter.
5.1.8 Set the apparatus in the Biosafety cabinet.
5.1.9 Ensure Operation is done in the safe working zone.
5.1.10 Ensure the adequate protective gear is worn during the operation such as gloves, cap, and mask.
5.2 Cleaning of Bio-safety Cabinet
5.2.1 Keep the Bio-safety cabinet interior and an exterior free from dust.5.2.2 Before starting work, clean with 70% Isopropyl alcohol (IPA) / Disinfectant using a non-fiber shedding cloth.
5.2.3 Inside cleaning of the Bio-safety cabinet is recommended in the following cases
5.2.4 Before starting any work in the cabinet.
5.2.5 After working in the cabinet.
5.2.6 Whenever there is a change of work program.
5.2.7 In the event of liquid spilling on the worktable.
5.2.8 Switch “ON” the Ultra Violet (UV) light for at least 15 minutes after handling the culture in Bio-Safety cabinet.
5.2.9 Before and after handling of culture, sanitize the outer surface of culture tube with disinfectant.
5.3 Shut - down procedure
5.3.1 Disconnect all utilities.5.3.2 Turn off the burner and close the gas tap.
5.3.3 Switch off the main motor.
5.3.4 Clean down the entire area with 70 % Isopropyl alcohol/ Disinfectant
5.4 Record the Details like burning of UV light, Velocity, Differential pressure, Cleaning, Instrument switched On/Off.
5.5 Monitoring of Bio-Safety cabinet
Monitoring of bio-safety cabinet can be done by following ways,5.5.1 Passive air sampling
5.5.1.1 Transfer the Petri plate into pass box (LAL test room to Incubator).
5.5.1.2 Enter the respective area as per the SOP for entry and exit.
5.5.1.3 Decontaminate the external surface of the Petri plates with the help of a sterile mop soaked in a filtered sporicidal agent.
5.5.1.4 Mark the Petri plates at the base with the following details with marker
Media lot no
Name of the location
Date of exposure
Exposed by
5.5.1.5 Place the Petri plates on the bench of bio-safety cabinet and remove the upper lid of the Petri plate and keep it in an inverted position.
5.5.1.6 Expose the media plate for 4 hours, After 4 hours of exposure, close the Petri plate with lid. Collect the Petri plate and transfer the plates into the incubator.
5.5.1.7 Incubate the exposed Petri plate in an inverted position in the incubator at 32.5 ± 2.5°C for 5 days.
5.5.1.8 After incubation observe and count the number of colonies on the colony counter or in the light source with the help of a marker.
5.5.1.9 Note down the observation
5.5.1.10 If the counts obtained are above the limits specified below investigate the results and take necessary actions as per SOP for
5.5.1.11 Frequency of Monitoring
Class A: Once in a day whenever there is activity.
5.5.1.12 Acceptance criteria
Action limit: Class A: 1 CFU / plate
5.5.2 Active air sampling
5.5.2.1 Transfer the required number of media cassette, plate, and accessories into pass box (LAL test room to Incubator).
5.5.2.2 Enter the respective area as per the SOP for entry and exit.
5.5.2.3 Decontaminate the external surface of the plates with the help of a sterile mop soaked in a filtered sporicidal agent.
5.5.2.4 Mark the Petri plates at the base with the following details with a marker,
Media lot no
Name of the location
Date of exposure
Exposed by
5.5.2.5 Place the media cassette/ Petri plate on the Air sampler and operate the instrument as per SOP for Air Sampler.
5.5.2.6 After completion of sampling collect all the plates and incubate in an inverted position in the incubator at 32.5 ± 2.5°C for 5 days.
5.5.2.7 After incubation observe and count the number of colonies on the colony counter or in the light source with the help of a marker.
5.5.2.8 Note down the observation
5.5.2.9 If the counts obtained are above the limits specified below investigate the results and take necessary actions as per SOP.
5.5.2.10 Frequency of Monitoring Class A: Once in a day
5.5.2.11 Acceptance criteria
Action limit Class A: 1 CFU / plate
5.5.3.1 Keep the non-viable particles counter in the pass box (LAL test room to Incubator).
5.5.3.2 Enter the respective area as per the SOP
5.5.3.3 Operate the sampler according to SOP for Air Sampler.
5.5.3.4 Perform the non-viable particle count at two different locations after completion of the activities.
5.5.3.5 Record the details of non-viable particles count.
5.5.3.6 Frequency Once in fifteen days.
6.0 ABBREVIATIONS
6.1 SOP - Standard operating procedure6.2 CFU - Colony forming units
6.3 UV - Ultraviolet
Also see: SOP for Cleaning and Operation of Bio-safety Cabinet
Get ready to use editable documents in MS-Word FormatView List
No comments:
Post a Comment
Please don't spam. Comments having links would not be published.